10.1.4 Maintenance
The output of an ultraviolet light deteriorates with burning age. The useful life of the light is approximately 5000
hours under specific test conditions.
The control system monitors UV light usage hours and when the 5000 hour point is reached. The UV light LED
indicator will blink fast along with the red LED alarm indicating the UV light should be replaced. To reset the timer,
perform the following:
•
Press and hold key the press UV light
–
UV light
–
UV light key sequentially and the reset will occur
removing the above alarm condition.
•
It is also possible to approximate the UV light usage by the following:
•
Press and hold key and UV light key together for three seconds. The
PresurFlow™
LED’s will then
indicate usage of approximately 450 hours per LED light from left to right.
Indicates approximately 450 hours used
Indicates approximately 900 Hours used
Indicates approximately 3600 hours used
Note:
Before testing with lamp off, the light may be cleaned with a lint-free cloth dampened with alcohol
or ammonia and water.
Summary of Contents for LabGard NU-540-300E
Page 7: ...OM0280 Page 7 of 63 Rev 3 July 2021...
Page 9: ...OM0280 Page 9 of 63 Rev 3 July 2021...
Page 10: ...OM0280 Page 10 of 63 Rev 3 July 2021...
Page 11: ...OM0280 Page 11 of 63 Rev 3 July 2021...
Page 12: ...OM0280 Page 12 of 63 Rev 3 July 2021...
Page 16: ...OM0280 Page 16 of 63 Rev 3 July 2021...
Page 20: ...OM0280 Page 20 of 63 Rev 3 July 2021...
Page 58: ...OM0280 Page 58 of 63 Rev 3 July 2021...
Page 59: ...OM0280 Page 59 of 63 Rev 3 July 2021...
Page 60: ...OM0280 Page 60 of 63 Rev 3 July 2021...
Page 61: ...OM0280 Page 61 of 63 Rev 3 July 2021...