OM0280
Page 40 of 63
Rev 3 July/2021
7.5.3.4 PresurFlow Alarm Verification
The PresurFlow Alarm setpoints are based on the calibration setpoint. Once the calibration setpoint is
entered, the Alarm setpoint offset pressure values will align from the calibration pressure value. The high or
low alarm setpoint can be verified by measuring inflow volume/velocity while adjusting blower up or down
within the Alarm Verification menu.
•
Press and hold
key, then press ↑ ↓ ↑ sequentially releasing the
key after the 3 key
sequence.
•
Note: If blower was off while entering into the Alarm Verification Menu, the low alarm limit will
immediately activate. Turn on blower; once airflow is above the low alarm limit, the alarm will turn
off.
•
LED segments will indicate blower speed percentage.
•
Press ↑ or ↓ key to raise or lower blower speed. Alarm is active so yellow and red LED’s will activate
if pressure reaches the low or high alarm limit.
•
Press
to exit (blower should turn off and not go through normal start up procedure).
7.5.3.5 Nite Care Calibration (See Section 8.2.2 to Activate the Nite Care Function)
The Nite Care mode is defaulted to operate the blower at approximately 600 rpm or a 14% duty cycle.
However, if desired the Nite Care blower speed can be adjusted higher or lower by performing the following:
•
Press and hold
key, then press hidden
–
blower -
↓ keys sequentially.
LED indicator above blower (green) key will blink fast.
Press ↑ or ↓ keys to adjust blower speed.
•
LED segments will indicate Nite Care blower speed percentage and active blower speed adjustment
o
Left end red LED indicates active blower speed adjust
The red LED will blink as soon as any adjustments are made and will continue to blink as the motor
rpm settles. Once the red LED stops blinking, the motor will run steady state at the new percentage.
Note: The red LED must be non-blinking to save or exit
o
Green LED’s indicate percentage on of scale (0
-100%)
o
Yellow LED’s indicate minimum (left/maximum (r
ight) blower speed has been achieved
Note: At any time during the process
•
Press hidden key to accept Nite Care blower speed calibration point
•
Press
key to abort and exit
Summary of Contents for LabGard NU-540-300E
Page 7: ...OM0280 Page 7 of 63 Rev 3 July 2021...
Page 9: ...OM0280 Page 9 of 63 Rev 3 July 2021...
Page 10: ...OM0280 Page 10 of 63 Rev 3 July 2021...
Page 11: ...OM0280 Page 11 of 63 Rev 3 July 2021...
Page 12: ...OM0280 Page 12 of 63 Rev 3 July 2021...
Page 16: ...OM0280 Page 16 of 63 Rev 3 July 2021...
Page 20: ...OM0280 Page 20 of 63 Rev 3 July 2021...
Page 58: ...OM0280 Page 58 of 63 Rev 3 July 2021...
Page 59: ...OM0280 Page 59 of 63 Rev 3 July 2021...
Page 60: ...OM0280 Page 60 of 63 Rev 3 July 2021...
Page 61: ...OM0280 Page 61 of 63 Rev 3 July 2021...