OM0280
Page 5 of 63
Rev 3 July/2021
LabGard® ES Energy Saver
Class II Laminar Flow
Biosafety Cabinet
Models
NU-540-300E/400E/500E/600E
Manufactured by:
NuAire, Inc. - Plymouth, Minnesota, U.S.A.
1.0
General Information
1.1 Description
The LabGard® ES Model NU-540E Laminar Flow Biosafety Cabinet (LFBSC) is a bench/table top model, optionally available
with a base support stand, for operation as a console model. The LabGard® ES model NU-540E utilizes an Energy Saver DC
ECM motor optimally determined forward curved fan for each model size/width to maximize both energy efficiency and
filter loading capacity. The Energy Saver ECM motor is controlled by the
Aeromax™
Control System to nominal airflow
setpoints via an integrated solid-state DC motor controller based on feedback algorithm to provide automatic
compensation (constant volume control) for both filter loading and line voltage variances. Airflow velocity is monitored by
the
PresurFlow™
system with a digital pressure sensor.
The Laminar Flow Biosafety Cabinet, (LFBSC) is a product resulting from the development of the "laminar flow" principle
(see Airflow Schematic) and the application of environmental controls as required in the field of biological research or
chemical containment. The LFBSC, when used with proper technique, is an effective primary engineering control in
obtaining the optimum product quality while reducing the potential for exposure of both product and personnel to airborne
biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation or product
operations, as prescribed by the Center for Disease Control (CDC) Atlanta, Georgia.
The NU-540E bench LFBSC meets the requirements of a Class II, since the cabinet conforms to the following requirements:
1. Maintains a minimum inflow velocity of 100 LFPM (.51mps) through the work access opening.
2. Has HEPA filtered downflow air that is mixed with the inflow air from a common exhaust plenum.
3. Discharges a percentage of air to the outside atmosphere after HEPA filtration.
4. Has all biologically contaminated ducts and plenums under negative pressure or surrounded by negative pressure.
Cabinets used for work with minute quantities of non-flammable or explosive volatile toxic chemicals and tracer amounts of
radionuclides required as an adjunct to microbiological studies must be exhausted through properly functioning exhaust
canopies.
Summary of Contents for LabGard NU-540-300E
Page 7: ...OM0280 Page 7 of 63 Rev 3 July 2021...
Page 9: ...OM0280 Page 9 of 63 Rev 3 July 2021...
Page 10: ...OM0280 Page 10 of 63 Rev 3 July 2021...
Page 11: ...OM0280 Page 11 of 63 Rev 3 July 2021...
Page 12: ...OM0280 Page 12 of 63 Rev 3 July 2021...
Page 16: ...OM0280 Page 16 of 63 Rev 3 July 2021...
Page 20: ...OM0280 Page 20 of 63 Rev 3 July 2021...
Page 58: ...OM0280 Page 58 of 63 Rev 3 July 2021...
Page 59: ...OM0280 Page 59 of 63 Rev 3 July 2021...
Page 60: ...OM0280 Page 60 of 63 Rev 3 July 2021...
Page 61: ...OM0280 Page 61 of 63 Rev 3 July 2021...