OM0280
Page 41 of 63
Rev 3 July/2021
Table 7.0
Recommended Measurement Methods for Cabinet Downflow & Inflow
A. Downflow Measurement
a. Recommended Instruments: TSI 8355 Thermo anemometer
b. Procedure:
Supply filter efflux is measured on a grid, in a horizontal plane 4 inches (102mm) above the bottom edge of the
window. No reading should be taken closer than 6 inches (152mm) from the inside perimeter.
c. Test Data - Inches (mm):
Model Size
Window Access Height
300E
10
6.25
(159)
13.542
(344)
20.834
(529)
28.125
(714)
400E
10
7.0
(178)
12.396
(315)
17.792
(452)
23.188
(589)
28.584
(726)
33.980
(863)
39.375
(1000)
500E
10
7.25
(184)
13.518
(343)
19.786
(503)
26.054
(662)
32.322
(821)
38.590
(980)
44.858
(1139)
51.125
(1229)
600E
10
7.5
(191)
15.410
(391)
23.320
(592)
31.230
(793)
39.140
(994)
47.050
(1195)
54.960
(1396)
62.870
(1597)
6.25
(159)
7.0
(178)
7.250
(184)
7.5
(191)
12.5
(318)
12.5
(318)
12.5
(318)
12.5
(318)
18.750
(476)
18.0
(457)
17.750
(451)
17.5
(445)
Number of Readings:
Average Velocity
ft./min.(m/s)
d. Acceptance Criteria:
1. Average downflow velocity =
55 to 65 fpm (.28 to .33 m/s)
2. Individual readings must be 20% or + 16fpm (+ 0.08m/s) whichever is greater (factory test)
or + 25% or + 16fpm (+ 0.08m/s) whichever is greater (field test) from the average downflow velocity.
B. Inflow Measurement
a. Recommended Instrument: Shortridge Flowhood ADM-870 or TSI 8355 Thermo anemometer.
b. Primary Procedure:
The primary procedure to determine inflow velocity uses a Direct Inflow Measurement (DIM) Instrument (i.e.
Shortridge flowhood).
The DIM Instrument can be used directly on the cabinet with NO CORRECTION FACTORS REQUIRED if operated in
the local density default mode.
NSF has tested the cabinet and established listed air velocities expressed in local density.
The DIM Instrument should also be duct taped to the cabinet to prevent any sneak air paths from occurring.
The DIM Instrument will read inflow volume (i.e. CFM).
Use the window access opening area to calculate inflow velocity.
Summary of Contents for LabGard NU-540-300E
Page 7: ...OM0280 Page 7 of 63 Rev 3 July 2021...
Page 9: ...OM0280 Page 9 of 63 Rev 3 July 2021...
Page 10: ...OM0280 Page 10 of 63 Rev 3 July 2021...
Page 11: ...OM0280 Page 11 of 63 Rev 3 July 2021...
Page 12: ...OM0280 Page 12 of 63 Rev 3 July 2021...
Page 16: ...OM0280 Page 16 of 63 Rev 3 July 2021...
Page 20: ...OM0280 Page 20 of 63 Rev 3 July 2021...
Page 58: ...OM0280 Page 58 of 63 Rev 3 July 2021...
Page 59: ...OM0280 Page 59 of 63 Rev 3 July 2021...
Page 60: ...OM0280 Page 60 of 63 Rev 3 July 2021...
Page 61: ...OM0280 Page 61 of 63 Rev 3 July 2021...