MODE SELECT
connectors (see also chapter 3.4)
All MODEs OF OPERATION can be selected by TTL signals connected to the rear panel
(see below), if the MODE OF OPERATION switch (
42
, Figure 1
)
is in EXT position. This
is very convenient when switching often between electroporation and recording, because
this can be done automatically by the data acquisition system using TTL signals.
(
7
) R
S
connector
BNC connector for remote control of the series resistance compensation circuit. A TTL HI
(+5 V) signal is connected here to activate the series resistance compensation circuit
remotely.
(
8
) R
EL
connector
BNC connector for remote control of the electrode resistance test. A TTL HI (+5 V) signal
can be connected here to select the electrode resistance test remotely.
(
9
) BR connector
BNC connector for remote control of the bridge mode. A TTL HI (+5 V) signal can be
connected here to select the bridge mode remotely.
(
10
) OFF connector
BNC connector to switch the ELC-03XS in OFF mode remotely with a TTL HI (+5 V)
signal.
(
11
) x10 MODE connector
BNC connector to switch the ELC-03XS to x10 mode of operation remotely (TTL HI
signal). Dependent on the signal level at
12
, this is CCx10 (signal at
12
= LOW) or VCx10
(signal at
12
= HI).
(
12
) VC / CC connector
BNC connector for remote control of the VC / CC mode of operation. A TTL signal can be
connected here to select the mode of operation remotely (HI = VC, LO = CC).
(
13
) GROUND connector
Banana plug providing internal ground (see below).
(
14
) CHASSIS connector
Banana plug providing mains ground (see below).
(
15
) FUSE holder
Holder for the line fuse. For changing the fuse rotate the holder counter-clockwise using a
screw driver.
(
16
) LINE SELECT switch
Switch for selecting the line voltage. Switch to the right for 230 V AC, to the left for 115 V
AC. The selected voltage is indicated on the switch.
Caution
: Before turning on the instrument, make sure that the correct line voltage is selected.