) REMOTE connector
BNC connector to attach a remote switch in active-low configuration for activating
the PENETRATION / BUZZ unit.
(
4
) MANUAL push button
Push button for activating the PENETRATION / BUZZ unit manually. The
DURATION is set by
48
and the mode by
2
.
Note
: The duration is dependent on the setting of the DURATION potentiometer, but
independent from how long the button is pushed or a REMOTE switch is pressed.
-12000
-8000
-4000
0
4000
8000
12000
time
potential (mV)
start
stop
Figure 2: BUZZ function of the ELC-03XS (CC Mode)
(
5
) FREQUENCY potentiometer
Potentiometer to set the FREQUENCY of pulses within a train in PULSE mode
(see also
2
). Range: 100 Hz to 2 kHz.
(
47
) AMPLITUDE potentiometer
Potentiometer for setting the AMPLITUDE of pulses in PULSE mode (see also
2
).
(
48
) DURATION potentiometer
Control to set the duration of the BUZZ (potentiometer, clockwise, range: ~1 ms to
~100 ms). It is effective in all modes; REMOTE controlled and when pushing the
MANUAL button.