BA-01X User Manual
_________________________________________________________________________________________________________________
___________________________________________________________________________
version 1.9
page 8
(1) POWER pressure switch
Switch to turn POWER on (switch pushed) or off (switch released).
(2) POTENTIAL / RESISTANCE display
Display for the recorded potential in mV (XXX mV) or the electrode
resistance in M
Ω
(XXX M
Ω
, i.e. 100 correspond to 100 M
Ω
), selected by
push button (31).
(3) M
Ω
LED
LED indicating that the unit of display (2) is M
Ω
.
(4) mV LED
LED indicating that the unit of display (2) is mV.
OSCILLATION SHUT-OFF unit
The OSCILLATION SHUT-OFF unit consists of (5) OSCILLATION SHUT-OFF
LED, (6) DISABLED/RESET switch and the (30) THRESHOLD potentiometer.
(5) OSCILLATION SHUT-OFF LED
Indicates whether the OSCILLATION SHUT-OFF circuit is active (LED: red) or
not (LED: green).
(6) DISABLED/RESET switch
Switch to DISABLE the OSCILLATION SHUT-OFF unit or RESET the circuit. RESET is
for resetting the circuit after previous activation. After resetting the OSCILLATION
SHUT-OFF unit is active again.
(30) THRESHOLD potentiometer
Control to set the activation THRESHOLD of the OSCILLATION SHUT-OFF circuit
(potentiometer, linear clockwise, range: 0-1200 mV).
Note: If the OSCILLATION SHUT-OFF unit is active, the output of the amplifier to the
headstage is disabled. Potential measurement works.