Rated output of the
transmitter [W]
150 kHz to 80 MHz 80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0,12
0,07
0,1
0,37
0,22
1
1,17
0,70
10
3,69
2,21
100
11,67
7,00
For transmitters whose maximum output rating is not specified in the above table, the recommended
separation distance d in metres (m) can be calculated using the equation found in the relevant column,
where P is the maximum output rating of the transmitter in watts (W) according to the transmitter
manufacturer‘s specification.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: This guideline may not be applicable in all cases. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Recommended separation distances between portable and mobile RF
telecommunications devices and the ProLift Cacero.
ProLift Cacero series hosts are intended for operation in an ELECTROMAGNETIC
ENVIRONMENT in which RF interference is monitored. The customer or user of the
ProLift Cacero hoist can help to avoid electromagnetic interference by maintaining
the minimum distance between portable and mobile RF telecommunications devices
(transmitters) and the ProLift Cacero hoist - depending on the output rating of the
communication device, as specified below.
Separation distance, depending on transmission frequency [m]
Recommended separation distances between portable and mobile RF
telecommunications devices and the medical equipment device or system - for
medical equipment devices and systems that are not life-supporting.
Summary of Contents for 960800