Your safety, the safety of your team, and of course that of your patients is very important to us. It is
therefore essential to bear the following information in mind:
Every use of the MD 11 different to the product description defined in chapter “Intended use and operation”,
causes risks for patients and trained personnel. If physical examinations and therapies are carried out
without use of the devices then the devices must be removed from the place of treatment. Avoid any
connection or close adjacency to other devices.
3.1
EMC Manufacturer's Declaration of Conformity
The use of (RF) Radio Frequency emitting devices and equipment as well as the occurrence of negative en-
vironmental factors in the close area of the MD 11 may cause unexpected or adverse operation. Use only
accessories and cables as specified in the product description. Further observe the EMC manufacturers
declaration of conformity.
3.2
Integrated peristaltic tube pump
The integrated peristaltic pump is used to cool tissue in order to prevent damage to the tissue. It may only
be operated with watery solutions, such as 0.9 % Sodium Chloride irrigation solution or “Ringer” solution.
Supplying medication using the integrated pump is expressly prohibited.
Important information
Autoclavable at 135°C for at least 5 min.
Do not use if the packaging is damaged
Suitable for thermal disinfection
Warning
Sterilized using ethylene oxide
Manufacturer
Observe the instructions for use
The device is designed for intermittent
duty operation at “1 min ON/3 min OFF”
at 4 cycles, after 15 min. brake.
Electrical and electronic devices that have
reached the end of their service life comprise
hazardous waste and may not be disposed of
together with household waste. Valid local
disposal regulations apply.
Type B applied part
Applied parts is the instrument
Symbol indicating the serial number with the
date of manufacture (year/month).
Do not reuse
Symbol indicating the order number.
Biohazard
Symbol indicating the lot number.
Motor
Date of expiry
Pedal Protective
ground
Protection against contact and continual
submerging.
Certified by the Canadian Standards
Association (CSA)
Date of manufacturing
CE symbol with notified body