© Nous Imaging, Inc.
FIRMM
®
Installation Guide - Siemens
Page 18
320-00201-DO-05
Revision 12
Released: 02/2022
Note:
Make sure you have followed the procedure above to properly power off the MRI Host
Computer and scanner before continuing.
In the diagram above, the existing internal Siemens network Ethernet cable
is removed from the
back of the MRI host computer internal network port and inserted into the provided 5-port
Ethernet switch. This cable leads to the Siemens Measurement and Reconstruction Subsystem
(MARS) in the equipment room.
The provided straight ethernet cable
is inserted between the provided 5-port Ethernet switch
and the Siemens internal network connection on the MRI host computer.
The provided right-angle Ethernet cable
is inserted between the switch and the FIRMM tablet
positioned on the desktop in view of the operator. Do not forget to plug the 5-port Ethernet switch
into AC power.