© Nous Imaging, Inc.
FIRMM
®
Installation Guide - Siemens
Page 16
320-00201-DO-05
Revision 12
Released: 02/2022
Press the ”
Add license
” button, enter the key as shown below and press the ”
Submit
” button when
it is entered. You will note that FIRMM will insert a “–“ between every fifth character automatically.
The About screen will display the correct state of the new license.
8
Connecting the Network
The provided 6ft CAT-6 cable (or user-supplied CAT-6 cable) and the provided 5-port Ethernet
switch may be used to provide an additional connection point from the FIRMM Tablet Computer to
the Siemens Host PC
internal
network.
The provided 5-port Ethernet switch will be inserted between the MARS in the equipment room
and the MRI Host Computer in the control room.
Note
:
It is important to shut down the Siemens MRI Host Computer and scanner using the “System –
End Session” and the MRI System Wall Interface prior to removing the original network connection
and installing the FIRMM.
Note
:
If your Siemens MRI Host Computer is connected to an uninterruptible AC power output it is
important that the FIRMM network switch also is plugged into an uninterruptible AC power source.
These are sometimes identified as, or referred too, as red outlets in a clinical installation.