Chapter Four: Installation and Operation
© 2018 Cutting Edge Optronics, Inc.
34
Patara I and II Laser User Manual
WARNING.
Do not set the temperature of SHG below air condensation temperature
(dew point) at laser head. Condensation on the SHG crystal can seriously damage the
crystal.
Check the eDrive
1.
Flip the power switch on the back panel of the eDrive to the
ON
or
-
position
2.
Make sure that the red
EMERGENCY
button is released. Turn on the eDrive
by turning the key to
ON
position and pressing the
POWER
button to power
up the eDrive.
3.
Check all the laser parameters in the eDrive Nitro. The menu designations will
help locate the settings. Use the following values for the Patara laser:
Menu 1
Menu 2
Menu 3
Parameter
Setting or Value
Channel Setup
Internal Trigger
Enabled
Channel Setup
Q-Switch
Q Switch (QS)
Enabled
Set Frequency
refer to test report
Set Window Width
5 µs
Set Q-Switch Power
100%
Channel Setup
Q-Switch
Set FPS Settings
FPS
Enabled
FPS Mode
Standard
FPS Delay
250 µs
Start Power
refer to test report
FPS Window Length
refer to test report
FPS Modulation Type
Exponential
PPK Open Offset
0
PPK Closed Offset
0
Channel Setup
Channel 1
Channel 1
Enabled
Set Current
refer to test report
Set Standby Current
0.0 A
Set Slew Rate
5 A/S
Slew Control
Enabled
Set Current Limit
30 A
Channel Setup
Channel 1
Fault Setup
Set Voltage Dropout
20 V
Set Current Tolerance
Enabled
Set Current Tolerance
4 A