6.9
Classification
Safety standard
IEC 60601-1
The type of protection against electric shock
Class
Ⅰ
equipment.
The degree of protection against electric shock
Type BF, CF applied parts
Electro-Magnetic Compatibility
Group I, Class A
Protection against ingress of water
IPX1
6.10
Default Alarming Values of All Parameters
Mode
Parameter
Adult
Infant
Neonate
Heart Rate
High
180bpm
200bpm
220bpm
Low
40bpm
50bpm
50bpm
Respiration
High
40bpm
50bpm
60bpm
Low
10bpm
10bpm
10bpm
Temperature
High
39
℃
39
℃
39
℃
Low
35
℃
35
℃
35
℃
Systolic
High
180mmHg
130mmHg
110mmHg
Low
60mmHg
50mmHg
50mmHg
Diastolic
High
120mmHg
90mmHg
90bpm
Low
50mmHg
40mmHg
30bpm
MAP
High
160mmHg
110mmHg
100mmHg
Low
50mmHg
40mmHg
30mmHg
SpO2
High
100%
100%
100%
Low
90%
85%
85%
Pulse Rate
High
180bpm
200bpm
220bpm
Low
40bpm
50bpm
50bpm
Alarm high and low limits: when the parameter value is higher than the high limit or lower than the low
limit, the system will alarm.
In the monitoring screen, press the Alarm key to activate the alarm timer. The time shows up on the
high limit left corner of the screen. When the alarm timer is activated, the system will activate the alarm
function when the specified time has passed.
Alarm timer has 4 options: 2 min, 5 min, 10 min, 20 min, or during alarm.