User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
11 -
Large-sized Adult Cuff
33cm ~ 47cm
17.4cm
Prior to use of the cuff, empty the cuff until there is no residual air inside it to ensure accurate
measurement.
When putting on the cuff, unveil and wrap it around the upper arm evenly to appropriate
tightness as shown below.
Figure 3.6 Cuff Position
When unplugging the cuff, hold the head of the connector and pull it out.
Safety Instructions for NIBP Monitoring
When taking the measure of an infant or neonate’s (less than 10 years old) blood pressure, do NOT
operate in the adult mode. The high inflation pressure may cause lesion or even body putrescence.
It is recommended to take the blood pressure measurement manually. Automatic measurement
should be used at the presence of a doctor/nurse.
NIBP monitoring is prohibited to those who have severe hemorrhagic tendency or with sickle cell
disease, or partial bleeding will appear.
Do NOT bind NIBP cuff on limbs with transfusion tube or intubations or skin lesion area, otherwise,
damages may be caused to the limbs.
Pay attention to the color and sensitivity of the limb when measuring NIBP; make sure the blood
circulation is not blocked. If blocked, the limb will discolor, please stop measuring or remove the cuff
to other positions. Doctor should examine this timely.
Do NOT twist the cuff tube or put heavy things on it.
3.3.3 To connect the SpO
2
SpO
2
probe is very delicate equipment. Please follow the steps and procedures in operating it. Failure to
operate it correctly can cause damage to the SpO
2
probe.
Operation procedure:
1. Connect the SpO
2
probe to the
right panel’s jack labeled”SpO2”.
When unplugging the probe, be
sure to hold the head of the connector and pull it out.
2. Insert one finger into the probe (index finger, middle finger or ring finger with proper nail length)
according to the finger mark on the probe, shown as below.