![North-Vision Tech Deluxe-100 User Manual Download Page 20](http://html.mh-extra.com/html/north-vision-tech/deluxe-100/deluxe-100_user-manual_1708130020.webp)
User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
12 -
Figure 3.7 Demonstration of SpO
2
Probe
The scenario above will cause inaccurate reading or no readings during SpO
2
measurement.
Safety Introductions for SpO
2
Monitoring
Continuous use of fingertip SpO
2
sensor may result in discomfort or pain, especially for those
patients with microcirculatory problem. It is recommended that the sensor should NOT be applied to
the same finger for over two hours.
SpO
2
measuring position must be examined more carefully for some special patient. Do NOT install
the SpO
2
sensor on the finger with edema or fragile tissue.
Do NOT put the SPO
2
sensor and pressure cuff on the same limb, otherwise the NIBP measuring will
affect SpO
2
measuring and cause the alarm error.
Do NOT use the damaged SpO
2
sensor.
The SpO
2
sensor can not be immerged into water, liquor or cleanser completely, because the sensor
has no capability of waterproofness.
Please do not allow the cable to be twisted or bended.
Please do not use nail polisher or other cosmetic product on the nail.
The finger should be of normal length.
3.3.4 TEMP Transducer Connection
Multi-parameter Patient Monitor has two TEMP transducers to measure different body temperature.
Connecting methods:
1. Attach the transducers to the patient firmly;
2. Connect them to “TEMP” on the right panel.
Note: When unplugging the probe, be sure to hold the head of the connector and pull it out.
3.3.5 Battery Installation
1. Make sure that monitor doesn’t connect to mains power supply and stays in switch-off status.
2. Open the battery cover, insert the battery into any slot of battery compartment, and pay attention to the
instruction of polarity direction in the compartment. Do not reverse the battery.
3. Move baffle plate with hand to fasten the battery.
4. Remove the battery cover. (According to your need, you can insert one more storage battery to prolong
using time.)