8 Maintenance
The glass surface where the light is emitted should be occasionally inspected for
potential contaminants and, if necessary, cleaned. Use a soft cloth soaked in
alcohol for this. Do not flood the lamp with cleaning liquid in the process.
The maintenance of the cooling system is primarily geared toward the used
refrigeration unit. When using decalcified water with an anti-algae additive, no
maintenance to the cooling circuit should be expected. However, if calcium and
algae deposits do occur, we recommend the following process for cleaning the
lamps:
1. Run a solution consisting of vinegar and decalcified water (Mixing ratio 1:1 n.
weight) through the lamp for 15 minutes
2. In the next step, run a pure vinegar solution through in the opposite direction for
15 minutes.
3. Next, run decalcified water through for 15 minutes.
4. Finally, run decalcified water through in the opposite direction for 15 minutes.
CAUTION!
Attention:
Do not use additives or cleaning solutions that contain salts or chlorine.
9 Accessories
The BlueCure lamps can be operated by using an adapter attachment with fiber
optics or liquid light conductors, which enable a localized exposure with high-
intensity UV light even on areas that are difficult to access.
Please contact us for more detailed information.