Norco Eco-Stim NC89481 Instruction Manual Download Page 20

Field strengths from fixed transmitters, such as base stations for radio  

(cellular/cordless) telephones and land mobile radios, amateur radio,  

AM and FM radio broadcast and TV broadcast cannot be predicted  

theoretically with accuracy. To assess the electromagnetic environment 

due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the 

device is used exceeds the applicable RF compliance level above, the 

device should be observed to verify normal operation. If abnormal  

performance is observed, additional measures may be necessary, such  

as reorienting or relocating the device.

Over the frequency range 150 kHz to 80 MHz, field strengths should be 

less than [Vi] V/m.

Test specifications for ENCLOSURE PORT IMMUNITY  

to RF wireless communications equipment (Table 9)

Test

frequency

(MHz)

Band

a)

(MHz)

Service

a)

Modulation

b)

Maximum 

Power 

(W)

Distance 

(m)

Immunity

Test Level 

(V/m)

385

380- 

390

TETRA 400

Pulse  

modulation

b)

 

18Hz

1.8

0.3

27

450

430- 

470

GMRS 460, 

FRS 460

FM

c)

 ±5kHz 

deviation 

1kHz sine

2

0.3

28

710

704- 

787

LTE Band 

13,17

Pulse  

modulation

b)

 

217Hz

0.2

0.3

9

745
780
810

800- 

960

GSM800/900, 

TETRA 800, 

iDEN 820, 

CDMA 850, 

LTE Band 5

Pulse  

modulation

b)

 

18Hz

2

0.3

28

870

930

1720

1700- 

1990

GSM1800; 

CDMA 1900; 

GSM 1900; 

DECT; LTE 

Band 1, 3, 4, 

25; UMTS

Pulse  

modulation

b)

 

217Hz

2

0.3

28

1845

1970

2450

2400- 

2570

Bluetooth, 

WLAN,  

802.11 b/g/n,  

RFID 2450,  

LTE Band 7 

Pulse  

modulation

b)

 

217Hz

2

0.3

28

5240

5100- 

5800

WLAN  

802.11 

a/n

Pulse  

modulation

b)

 

217Hz

0.2

0.3

9

5500
5785

NOTE:

 If it is necessary to achieve the IMMUNITY TEST LEVEL, the distance  

between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM 

may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.

a) 

For some services, only the uplink frequencies are included.

b) 

The carrier shall be modulated using a 50% duty cycle square wave signal. 

c)

  As an alternative to FM modulation, 50% pulse modulation at 18Hz may  

be used because it does not represents actual modulation.

39

38

Summary of Contents for Eco-Stim NC89481

Page 1: ...Instruction Manual Be sure to read this instruction manual before operating device Keep manual in a safe place NC89481 Eco Stim TENS EMS...

Page 2: ...hat the device complies with the following normative documents IEC60601 1 IEC60601 1 2 IEC60601 1 11 IEC60601 2 10 IEC62304 ISO10993 5 ISO10993 10 ISO10993 1 ISO14971 QUICK REFERENCE GUIDE 4 1 FOREWOR...

Page 3: ...ent Press ON OFF M button to stop treatment and return to the standby mode QUICK REFERENCE GUIDE Please read the instruction manual completely before attempting to use this device Steps 1 3 should be...

Page 4: ...ds us that something is wrong without which abnormal conditions may go undetected causing damage or injury to vital parts of our bodies 1 FOREWARD Pain does not occur until encoded messages travel to...

Page 5: ...cose veins etc 2 SAFETY INFORMATION 4 Electrode placements must be avoided in the carotid sinus area anterior neck or transcerebrally through the head 5 This device should not be used in overly enerva...

Page 6: ...act with each other 19 Keep the stimulator out of reach of children 20 Consult your doctor if you have any questions 21 Discontinue use and do not increase the intensity level if you feel discomfort d...

Page 7: ...proper use may be dangerous 15 Rare cases of skin irritation may occur at the site of the electrode placement following long term application 16 Do not use this device in the presence of other equipme...

Page 8: ...ing mode press the button to unlock the keys In setting mode press the button to decrease the corresponding data for the pulse rate pulse width or treatment time 5 button In standby or treatment mode...

Page 9: ...ams P W Pulse Width 100 300 s P R Pulse Rate 2 100Hz Hz vibration per second Treatment time 5 90 minutes Number of programs 8 programs P W Pulse Width 100 300 s P R Pulse Rate 2 100Hz Hz vibration per...

Page 10: ...the connector into the port on the top of the device Ensure the electrode wires are inserted correctly The device has two output ports controlled by Channel A and Channel B at the top of the unit You...

Page 11: ...ice or the appropriate channel first in order to avoid unwanted irritation 5 It is recommended that at minimum 1 97 x 1 97 self adhesive square electrodes are used at the treatment area 6 Never remove...

Page 12: ...1 2 2 1 1 2 1 2 1 2 1 2 23 22 Arm Hand Back Abdomen Hip Leg Foot Foot Joint knee Joint elbow Joint ankle Joint wrist Neck Shoulder Position of electrode placement under EMS programs...

Page 13: ...isplays as follows 6 3 Select treatment program Based on your needs press P button to select the treatment program The LCD displays as follows 6 INSTRUCTIONS FOR USE 6 4 Set program parameter Press an...

Page 14: ...dby mode 6 5 Start treatment Press the button of CH1 to increase the channel 1 intensity press the button of CH2 to increase the channel 2 intensity The LCD displays as follows 6 6 Adjust the output i...

Page 15: ...nd contact a physician right away 3 Batteries must not be charged dismantled or short circuited 4 Protect batteries from excess heat Take the batteries out of the device if they are spent or you will...

Page 16: ...hile the device is still on 3 Only use the electrode pads recommended by the manufacturer Use of inferior electrode products could result in injuries to the user 6 11 Where do I attach electrode pads...

Page 17: ...before use 4 Check your allergic history Change the electrode location or shorten the treatment time If your skin is over sensitive stop the treatment or see a doctor 33 32 4 Clean the surface of the...

Page 18: ...isconnect the lead wires from the electrodes 3 Place the electrodes onto the plastic film and store in the sealed package 4 Wrap the lead wires and store in the sealed package 9 2 Storing the Unit 1 P...

Page 19: ...f a typical location in typical commercial or hospital environment NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic immun...

Page 20: ...r W Distance m Immunity TestLevel V m 385 380 390 TETRA 400 Pulse modulationb 18Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS 460 FMc 5kHz deviation 1kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modula...

Page 21: ...or invoice 2 Repairs under warranty should be in the warranty period for the device 3 The following cases are excluded under the warranty All damages that arise due to improper operation e g nonobserv...

Page 22: ...43 42 USER NOTES...

Page 23: ...orth Coast Medical Inc www ncmedical com Manufactured for North Coast Medical Inc 780 Jarvis Drive Suite 100 Morgan Hill CA 95037 U S A Made in China REV89481i_0521 Instruction Manual Eco Stim TENS EM...

Reviews: