Treatment
mode
Program
NO.
Treatment time
(min)
Frequency
(Hz)
Pulse width
(us)
Type
PU1
Default: 30
Adjustable:
(5-90)
Default: 50
Adjustable:
(2-100)
Default: 180
Adjustable:
(100-300)
Con.
TENS
PU2
Default: 30
Adjustable:
(5-90)
Default: 60
Adjustable:
(2-100)
Default: 160-260
Adjustable:
(100-300)
PWM
PU3
Default: 30
Adjustable:
(5-90)
Default: 60
Adjustable:
(2-100)
Default: 260
Adjustable:
(100-300)
IM
PU1
Default: 30
Adjustable:
(5-90)
Default: 5
Adjustable:
(2-100)
Default: 300
Adjustable:
(100-300)
Con.
EMS
PU2
Default: 30
Adjustable:
(5-90)
Default: 60
Adjustable:
(20-100)
Default: 200
Adjustable:
(100-300)
SY
PU3
Default: 30
Adjustable:
(5-90)
Default: 70
Adjustable:
(20-100)
Default: 200
Adjustable:
(100-300)
AL
27
26
4) Press [ON/OFF/M] button to return to the standby mode.
6.5 Start treatment
Press the [+] button of CH1 to increase the channel 1 intensity,
press the [+] button of CH2 to increase the channel 2 intensity.
The LCD displays as follows:
6.6 Adjust the output intensity
Place the electrodes on the body parts, press the [+] button to
increase output intensity. Intensity will be increased to a higher
level after each press. The device has 40 levels of output intensity.
Adjust the intensity to a level you feel comfortable. The level of
output intensity will be shown on the LCD:
If the intensity is too strong, you can press [-] button to decrease
intensity to a lower level. When the output intensity of both
channels decreases to zero, the stimulator will return to the
standby mode. The LCD displays as follows:
Caution
If you feel or become uncomfortable, reduce the stimulation
intensity to a more comfortable level and consult with your
medical practitioner if problems persist.
Summary of Contents for Eco-Stim NC89481