14 Appendix A: Goniometer Specifications
(Refer to figure above)
SG65
SG75
SG110
SG110/
A
SG150 SG150/B
Q110
Q150
F35
Number of
Channels
2
2
2
2
2
2
1
1
1
Dimensions mm
A. Maximum
65
75
110
110
150
150
110
170
35
A. Minimum
30
35
70
70
100
50
70
115
30
B.
55
60
60
60
70
120
60
70
18
C.
18
18
18
18
18
18
18
18
8
D
54
54
54
18
54
54
54
54
15
E
20
20
20
54
20
20
20
20
8
Weight (g)
22
23
23
29
25
27
22
23
8
Minimum
permissible
bend rad. (mm)
18
18
18
18
18
18
18
18
3
Measuring Range
150
150
150
150
150
150
150
150
150
Crosstalk
1
5%
5%
5%
5%
5%
5%
N/A
N/A
N/A
Transducer type
strain gauge
Life
2
600,000 cycles minimum
Accuracy
2
measured over 90
from neutral position
Repeatability
better than
1
Analogue resolution
infinite
Operating temperature range:
+ 0
C to +40
C
Storage temperature range:
-20
C to +50
C
Operating humidity range:
30% to 75%
Storage humidity range:
30% to 75%
Atmospheric pressure range:
operation:
700hPa to 1060hPa
storage:
500hPa to 1060hPa
15 Appendix B: Application Diagrams
Available Goniometer Sensor Types:
1
Specification of crosstalk for all Biometrics twin axis SG series goniometers is measured over
60
.
i.e. if a joint is moved through 60
from the neutral position in one plane without movement in the orthogonal
plane, then the sensor output in the orthogonal plane may change by a maximum
3
.
2
Life test results have been collected by cycling the sensors through movements that would happen
during everyday use. For example, placing a sensor on an adult elbow and moving from the neutral position to
maximum flexion and back to the neutral position, the unit will function for a minimum of 600,000 cycles.
B
A
D
C
E