2D Goniometer SmartLead User Manual
17
(Rev D January 2020)
individually and carefully making sure that the minimum permissible bend radius is
observed, particularly where the measuring element enters the endblocks.
When cleaning or disinfecting goniometers, remove them from the Ultium SmartLead
to prevent liquid from leaking into the Ultium SmartLead and Sensor.
When using SG series sensors, care should be taken during mounting so that the
measuring element always forms a “simple” bend shape.
If an “oxbow” shape occurs
in the element, this will reduce accuracy.
This results when the sensor is
compressed too much. This can damage
the sensor. Do NOT mount the sensor in
this way.
When using F series goniometers, the sensor should not be bent more than
20
in
the y-y plane, otherwise reduced life or failure will result.
Single Axis Goniometer F35
POSITION 1
L 1
!
!
!