34
Care and Maintenance
Care and Maintenance
Wipe the device with a soft cloth dampened with a mild detergent or 10% bleach solution. Do not use
undiluted bleach or any cleaning solution other than those recommended here, as permanent damage
could result. Dry with a soft cloth, or allow to air dry.
Clean the device separately from the sensors. For instructions on cleaning pulse oximeter sensors, refer
to the respective sensor instructions for use.
Maintenance
The 9847 requires no routine calibration or maintenance, other than battery replacement.
The Oxitest
Plus7
by Datrend Systems, Inc. can be used to verify operation of the pulse oximeter.
Cleaning the CO
2
Sensor
The CO
2
sensor is protected against splashing water.
Cleaning the CO
2
Sensor
Clean the Model 9840SA CO
2
Sensor with a soft cloth dampened with isopropyl alcohol. Do not pour or
spray any liquids onto the Model 9840SA CO
2
Sensor. Allow the Model 9840SA CO
2
Sensor to dry
thoroughly before reusing.
Returning the CO
2
Sensor for Service
If the Model 9840SA CO
2
Sensor must be returned to Nonin for service, the product should be free of
any contaminants, and sterilization may be required. Contact Nonin’s Technical Service department for
shipping instructions.
CAUTION:
Do not place the Model 9847 in liquid or clean it with agents containing
ammonium chloride, isopropyl alcohol, or products that are not listed in this User’s
Guide.
CAUTION:
Do not use caustic or abrasive cleaning agents on the device or the
sensors.
CAUTION:
Do not autoclave or immerse the device or sensors in liquid. Do not
expose the device or components to excessive moisture or liquids.
CAUTION:
Do not immerse the CO
2
sensor in liquid, and do not use caustic or
abrasive cleaning agents on the CO
2
sensor.
!
!
!
!