
16
Using the 9847 Pulse Oximeter
Carbon Dioxide Detector
Verify:
• that the airway adapter tube is properly attached to the CO
2
sensor,
• the CO
2
sensor is properly connected to the device,
• the lower bar of the CO
2
display is lit,
• the CO
2
sensor light is blinking, and
• the CO
2
bar graph segments indicate change in CO
2
level.
Attach the CO
2
sensor/airway adapter tube to the patient’s endotracheal tube. The CO
2
detector will reflect
values and breath beeps for each breath (if the breath beep sound volume is not turned off).
Medium priority (equipment) alarm will occur (if the alarms are not disabled), when:
• the CO
2
sensor is disconnected,
• the airway adapter tube is removed from the CO
2
sensor,
• the light path is blocked, or
• a CO
2
sensor failure occurs.
High priority patient alarms override medium priority (equipment). The third and sixth
segments of the CO
2
bar graph will be illuminated, indicating a CO
2
sensor fault.
Setup
All functions of the 9847 are controlled by buttons found on the keypad on
the front of the device.
Press the On/Standby button to turn the device on or to enter Standby
mode.
Setup Mode
Setup mode is used to control the audible and visible patient alarm limits,
the breath beep pitch, the calendar, and the internal time-of-day clock. In
setup mode, the alarm limit review button is used to make the menu
selections, and the advance button is used to change the value.
Entering Setup Mode
1. With the unit in Standby mode, press and hold the alarm limit review
button while pressing and then releasing the On/Standby button.
2. Release the alarm limit review button when “
888 888
” is displayed on the %SpO
2
and pulse
rate displays. Three brief beeps will sound, and “
ALr
” and the “
dFt
” will appear in the %SpO
2
and pulse rate displays.
NOTE:
Entering setup mode with audible alarms disabled:
1. Press and hold the alarm limit review and audible alarm disable buttons while pressing and
releasing the On/Standby button;
2. Release the alarm limit review and audible alarm disable buttons; and the display will appear
as described in #2 above; the audible alarm limit indicator will illuminate steadily.
ΔΔ
CO
CO2
m
m
H
g
k
P
a
75
30
50
2
6
10
20
10
4
7
0.3
0.8
1.3
2.6
On / Standby
Advance /
Breath Beep
Volume
Control
Audible
Alarm Disable
Alarm Limit
Review
Audible Alarm
Disabled
Indicator
(Yellow)
Keypad
Low Battery
(LED)
Indicator