35
Electrical Connections
6. Set the address bus select switches to a unique address for the SELECT and
ERD as shown in
.
Caution:
All modules on the External bus must use a unique address. A bus con
-
flict and the system communication failure occurs when two or more modules are
set to the same address.
Figure 19
Setting External bus addresses - single ERD system
7. Connect AC mains power to the ERD.
8. Turn ON the laser and ERD. Check that the SELECT is available in the Device
Selector panel of CONTROL (see
)
Dual ERD system
For basic operation, connect the External bus from the SuperK laser to the
SELECT and to both External RF Drivers (ERDs) where the bus is looped back
using the included bus defeater on the last module in the bus chain. The RF
coaxial BNC port of ERD 1 and ERD 2 must be connected to the RF1 and RF 2
BNC ports respectively on the SELECT. Follow
to make the
connections.
Procedure 3
Electrical connections for a SELECT with dual ERDs
1. Disconnect power from the laser.
Caution:
If the laser is powered ON when connecting the external bus cable,
DAMAGE can occur to both the laser and the SELECT.
2. Using an External bus cable, connect the
External bus
connector of the
SuperK laser to the
Bus Input
) of the SELECT.
External RF Driver:
Bus address = 2
SELECT module:
Bus address = 0
Summary of Contents for A203-000-001
Page 1: ...SuperK SELECT PRODUCT GUIDE Multi Channel Tunable Filter...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 30: ...Status LEDs 30...
Page 56: ...Control panel 56...
Page 59: ...59 Figure 44 Mechanical dimensions External RF Driver...
Page 60: ...60...
Page 62: ...Support Contact Details 62 2...
Page 68: ...Installing CONTROL 68...
Page 77: ...SuperK SELECT Product Description Revision 1 00 10 2020 W 10456...