Electrical Connections
36
Figure 20
Connecting a SELECT - dual ERD
3. Using the other External bus cable, connect the
Bus Through
connector of
the SELECT (item C of
) to the
Bus Input
connector of ERD 1 (item B
).
Figure 21
Connecting dual ERDs
4. Connect the
Bus Through
connector of ERD 1 to the
Bus Input
connector of
).
5. Place the bus defeater (
) onto the
Bus Through connector
of the
ERD
(item E of
).
Note:
Ensure to terminate the External bus with an NKT Photonics bus defeater,
Without the bus defeater, the interlock circuit remains open disabling laser emis
-
sion.
6. Using an RF coaxial cable, connect the
RF
BNC port of the ERD 1 (item A of
) to the
RF 1
).
7. Using the other RF coaxial cable, connect the
RF
BNC port of the ERD 2 (item
B of
) to the
RF 2
).
A B
C
A: RF2
from RF
out ERD2
B: RF1
from RF
out ERD1
C: Bus Through to External BUS
ERD1
D: Bus Input from External BUS
SuperK laser
D
A B
C
D
A: RF
ERD1
to RF1
Select
B: RF
ERD2
to RF2
Select
C: External bus
from Select
D: External bus
ERD1 to ERD2
E: Bus defeater
ERD2 bus thr.
E
ERD 1
ERD 2
Summary of Contents for A203-000-001
Page 1: ...SuperK SELECT PRODUCT GUIDE Multi Channel Tunable Filter...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 30: ...Status LEDs 30...
Page 56: ...Control panel 56...
Page 59: ...59 Figure 44 Mechanical dimensions External RF Driver...
Page 60: ...60...
Page 62: ...Support Contact Details 62 2...
Page 68: ...Installing CONTROL 68...
Page 77: ...SuperK SELECT Product Description Revision 1 00 10 2020 W 10456...