Thermo Scientific AquaSensors
™
AV88 Universal Analyzer User Guide
53
ALARM Function
Edit Range
Default Setting
Set Phase
High or Low
High
Set Setpoint
Min to Max scale
Highest reading
Set Deadband
0 to 14 pH
0 pH
Set Off-Delay
0 to 999 seconds
0 seconds
Set On-Delay
0 to 999 seconds
0 seconds
If WASH is selected, the following activation options can be programmed in the ACTIVATION menu.
ALARM Function
Edit Range
Default Setting
Set Interval
0.0 to 999.9 minutes
0.0 minutes
Set Duration
0 to 999 seconds
10 seconds
Set Off Delay
0 to 999 seconds
0 seconds
Navigation from the Measure Menu:
1.
Press MENU to display the Main Menu.
2.
Scroll with the UP/DOWN arrows to view RELAY, RELAY A or RELAY B.
3.
Press ENTER to select the desired relay.
4.
Scroll with the UP/DOWN arrows to view FUNCTION
5.
Press ENTER to select Function and use the UP/DOWN arrows to select Alarm, Control or Wash.
6.
Press ENTER to select Alarm, Control or Wash. Press ESC to leave the setting unchanged.
7.
Scroll with the UP/DOWN arrows to SET PARAMETER (unless Wash was selected).
8.
Press ENTER to select SET PARAMETER and use the Up/DOWN arrows to select SENSOR or
TEMPERATURE to drive the relay.
9.
Press ENTER to select Sensor or Temperature. Press ESC to leave the setting unchanged.
10.
Scroll with the UP/DOWN arrows to SET ACTIVATION.
11.
Press ENTER to select SET ACTIVATION and use the UP/DOWN arrows to scroll the list of activation
options. Select an option with the ENTER key. Edit the selected option with the UP/DOWN arrows.
Press ENTER to select the edited option or press ESC to leave the option unchanged.