Be sure the sensor has been properly cleaned before calibration. Initiate the 1-point sample process by
pressing ENTER.
The sensor value is dynamically displayed at this point in the calibration procedure as shown in Figure 4.6.
When the reading has stabilized, press ENTER to initiate the calibration.
Figure 4.6: Dynamic display of the sensor value during calibration.
When the user presses enter, the current sensor value is captured and shown in the edit field of the next screen
as shown in Figure 4.7.
Figure 4.7: The calibration edit screen.
1-Point Sample?
( 991.3)
S/cm
MENU
ESC
ENTER
1-Point Sample
991.3
S/cm
MENU
ESC
ENTER
1-Point Sample
Sample ready?
MENU
ESC
ENTER