4.
CALIBRATE MENUS
Calibration methods supported in the AV88 are automatically available depending on the personality module
installed in the AV88. The supported calibration functions are listed below:
Calibration Type
Function
Sensors Supported
1-Point Sample
Set measured reading to desired reading.
Changes offset. Best done at high end of range.
All
2-Point Buffer
Automatically sets measured readings to closest
buffer standard. Two points separated by more
than 2 pH units. Buffer standards are selected in
the configuration menu. Changes span and offset.
pH
Zero
Set measured reading to zero. Measured reading
must be close to zero.
Conductivity, Turbidity,
Dissolved Oxygen, Ozone.
Air
Adjust measured reading to calculated reading of
oxygen in air with pressure and salinity
configuration settings as well as temperature
reading. Assumes sensor is placed in air.
Dissolved Oxygen
Temperature 1-Point Set measured reading to desired reading.
Changes offset. Best done at high end of range.
All
Exit
Returns to the measure screen
Calibration functions that are not supported in the AV88 but are supported with the Thermo Scientific
AquaSensors remote protocol include:
2-Point Sample
Sets measured reading to desired reading for
two points separated by more than 2 pH
units. Changes span and offset.
NOT supported in AV88 local menu.
Accessed through digital network.
1-Point Buffer
Automatically sets measured pH reading to
closest buffer standard point. Buffer
standards are selected in the configuration
menu. Changes offset.
NOT supported in AV88 local menu.
Accessed through digital network.
Calibration choices vary depending on which personality module is installed in the AV88.