5.2.5
QUICKSET
Suggested for simple applications
Quick programming of the 4 basic parameters (like with programming without display module but) aided by images on the display. Measured values can only be
displayed in percentage.
QUICKSET can only be used in (%)(DEFAULT) mode (See
P01
in Full Parameter Programming 5.2.6).
Keys
Function
E
+
(min. 3 s)
Enter/Quit QUICKSET programming mode
+
LEARNING function
,
,
Adjustment of digit value (increase, decrease) moving blinking
E
Saving displayed value and step to the next screen
+
Reload value before modification (CANCEL)
E
+
(min 3 s)
Return to DEFAULT
+
Display DEFAULT
Screens
Actions
Assignment of 4 mA to the minimum level, %
Fill the tank to the level required. After entering the Quickset programming mode, the learning function can be started with the
double key pressing
+
. During learning the image „Store” is displayed.
The parameter value represents filling in %.
“Indirect” assignment of 4 mA to the minimum level with partially filled tank:
Following learning enter the momentary % value of level with the
buttons.
The output current will be set in accordance % value.