BKI16ATEX0010X
cbr2052a0600p_05
33/36
7.
ERROR
CODES
Error
code
Error description
What to do
Overwriting
indication
LED-Blinking
Current output
1
Memory failure
Contact service
Yes
Continuous
22 mA
2
Capacitance to be measured is too high or
damaged probe insulation
(Level over the upper limit)
Check adjustment and installation conditions.
Check the probe.
Repeat programming (teaching)!
Yes 2
Programmed
3
Hardware failure
(EEPROM communication failure)
Contact service
Yes
Continuous
22 mA
4
Display overflow
Check adjustment
Yes
-
No influence
5
Capacitance to be measured is too small
or no signal from the probe (short-circuit).
Check installation conditions
2
Programmed
6
Measuring at the limit of reliability.
No legible signal
(Noise, Electromagnetic disturbance)
Eliminate source of disturbance
Change the position of the probe or relocate it.
Yes 2
Programmed
12
Linearization failure: L(1) and L(2) zero
(no valid data pair)
See „Linearization”
Yes
3
22 mA
13
Linearization table failure: two L(i) in the
table
See „v Linearization”
Yes
3
22 mA
14
Linearization table failure: r(i) values are
not increasing continuously.
See „Linearization”
Yes
3
22 mA
15
Linearization table failure: no value
assigned to the measured one
See „Linearization”
Yes
3
22 mA
16
Incorrect checksum
Check programming. Change any parameter and
recheck the checksum.
Should the same error come out contact service!
Yes 3 22
mA