TRUE METRIX
®
PRO System Comprehensive Resource Guide (GDH-FAD Enzyme)
Quality Control Testing 21
11. Compare result to Control Test range printed on
test strip vial label for level of control solution
you are using. If result is in range, system can
be used for testing blood. If result does not fall
within range, repeat test using a new test strip.
Note: Control Test result shows the Control Symbol in
the Display.
If Control Test result is outside range,
test again. If result is still outside range,
the system should not be used for testing
blood. Call for assistance.
12. After result is shown, Strip Release Button
flashes. Hold meter with test strip pointing
down. Press Strip Release Button to release and
discard test strip in appropriate container. Meter
turns off.
Note: Removing test strip before result displays
cancels the test. An error message appears
and the result is not stored in Memory. Retest
with a new test strip and do not remove before
result is displayed.
Quality Control Testing
cont.
Control Symbol
Test Strip
Vial Label
Control Bottle
Label