NIPPON Genetics EUROPE – www.nippongenetics.eu
32
13.
A height of the tip ejector can be adjusted
so that it can match the conditions of how
the tip used was connected.
Use a small minus screwdriver.
By turning the adjustment screw in a
counter-clockwise direction, the tip ejector
can be lowered toward the tip side.
14.
When cleaning the outside of the pipette or the tip holder, remove dirt by using a cloth dampened
with 60% isopropyl, 70% ethanol or pH-neutral detergent, and wipe it using a lint-free and dry soft
cloth. Also, because the ejector can easily be removed, the tip holder can easily be cleaned.
Regularly clean the tip holder about once every half a year to a year, depending on the frequency
used. Regularly confirm the pipette accuracy.
Maintenance
When inspecting the pipette or managing pipette precision, the following device can be used. (Refer
to “17. LIST OF ITEMS SOLD SEPARATELY
(
DISPOSABLE ITEMS
)
”)
- Leak tester (Sold separately: AD-1690):
Leakage within the pipette can be easily checked.
- Pipette accuracy tester (Sold separately: BM series, AD-4212B-PT, FX-300i-PT):
By using an electronic balance and specified software “WinCT-pipette”, the actual dispensed amount
of the pipette can be confirmed.
Expanded
Ejector
Tip holder
Adjustment screw