Program setting mode
The nine programs can be saved in the program memory built into the pipette (PROG 01 to 09). By
saving a frequently used mode or volume for operation beforehand, these setting can easily be read
out from the next use. Select and set the mode or volume to be saved before saving the program
setting.
Saving the program setting
Set the pipette to your preferred settings.
Example: When saving the AUTO mode, dispensing volume 200
μ
L, buzzer ON and blowout ON
(The display is example.)
[ 1 ] Press and hold the Down key to enter
program setting saving mode.
[ 2 ] Press the Up or Down key to select program
number to be save from 1 to 9.
[ 3 ] Press the Enter key to save the setting.
Once these have been saved a buzzer will sound once.
Reading out the program setting
The set program can be read out at system setting mode (SYS).
(Refer to
“9-4 system setting mode (SYS)” for details)
10-6
Reverse operation (Dispensing liquid that tends to remain in
the tip)
When you would like to accurately dispense a highly viscous liquid that has a tendency to remain in
the tip, we recommend using the reverse operation. By aspirating a large amount of the liquid
beforehand, the reverse operation enables the correction of the amount of liquid remaining in the tip.
To enable reverse operation, set the setting of
at system setting mode (SYS).
(Refer to “9-4 System setting mode (SYS)”.)