Chapter 5
Assembling
79
Connecting signal cables
Connect the signal cable of the intelligent nosepiece to the
connector “NP” on the motorized focus unit or the relay
box. (See “[4] Connect the signal cables.”)
3
Attach the zooming body.
Attach the SMZ18 or SMZ25 Zooming Body to the intelligent nosepiece or focus mount adapter.
Both zooming bodies are attached to the intelligent nosepiece or focus mount adapter in the same manner.
The following figure shows an example of attaching the SMZ18 Zooming Body to the intelligent nosepiece.
When attaching to the focus mount adapter
Step (1) is not necessary when attaching the zooming body to the focus mount adapter. Proceed to Step (2).
(1) Detach the objective attachment adapter (dovetail
and ring) attached under the zooming body. Loosen
the three M3 hexagon socket head bolts using a hex
wrench (nominal designation: 2.5).
1
0.75
.
Detaching the objective attachment adapter
(2) Completely loosen the two M5 hexagon socket head
bolts for fixing the zoom body on the right side of the
zooming body (as viewed from the front) using a hex
driver (nominal designation: 4).
(3) Drop and fit the dovetail groove on the rear side of
the zooming body onto the dovetail of the intelligent
nosepiece mount.
Signal cable
Summary of Contents for SMZ18
Page 1: ... M608EN01 Research Stereo Microscope SMZ25 SMZ18 Instruction Manual M608 E 13 5 NF 1 ...
Page 2: ......
Page 20: ...Chapter 1 Components 6 ...
Page 32: ...Chapter 2 Microscopy Procedures 18 ...
Page 84: ...Chapter 4 Functions and Operations of the Devices 70 ...
Page 114: ...Chapter 6 Troubleshooting 100 ...
Page 118: ...Chapter 7 Maintenance and Storage 104 ...
Page 119: ...105 Chapter 8 Specifications ...