Chapter 3
Specific Operations
22
3
Switching the Optical Path of the Tube
For observing a specimen using the eyepiece, switch the optical
path to the binocular part. For imaging using the DS camera head,
switch the optical path to the vertical tube. Use the optical path
switching lever of the tube to switch the optical path. Push or pull
the lever to make the desired selection. Be sure the lever is fully
pushed in or pulled out. Do not leave the lever stopped partway.
Switching the optical path
Distribution of Light with Optical Path Switching (%)
Tube
Lever position
Binocular part
Vertical tube*
Left
Right
P2-TERG 100 Trinocular Tilting
Tube
Pulled out
100
100
0
Pushed in
100
0
100
P2-TERG 50 Trinocular Tilting Tube
Pulled out
100
100
0
Pushed in
100
50
50
P2-TL100 Trinocular Tube L
Pulled out
100
100
0
Pushed in
100
0
100
* The light amount of the left eye is always 100%. The light from the optical path of the right eye is diverged to the vertical tube.