Table specifications
Dimension
Details
Width
887 mm (35 inches) (not including wheel position)
Depth
600 mm (24 inches) (not including wheel position)
Table height range
725 mm to 1205 mm (29 inches to 47 inches)
Monitor Support Arm radius
1.
514 mm (20 inches)
Monitor mount type
VESA MIS-D, 75, C (hole pattern: 75 mm x 75 mm)
Monitor Support Arm load
3 kg (6.6 lbs) to 5 kg (11 lbs)
Table weight
53 kg (116.8 lbs)
Maximum load on table
65 kg (143.3 lbs)
Total mass
118 kg (260.1 lbs)
The table provides wheelchair accessibility.
Environmental considerations (ambient)
Do
NOT
operate the table in an environment that exceeds the environmental limits.
Status
Temperature
Relative Humidity
(Non-condensing)
Atmospheric
Pressure
Normal use
+10°C to +40°C
75% (maximum)
70kPa to 106kPa
1. The radius is the maximum distance that the Monitor Support Arm extends beyond the width of the table.
Guidance and manufacturer declaration – electromagnetic
immunity
Immunity test
IEC60601 compliance
test level
Electromagnetic environment-guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power supply
lines
±1 kV for input / output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC61000-4-5
±1 kV line to line
±2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
lines
IEC61000-4-11
<5% U
T
(>95%dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the AC mains voltage prior to the application of the test level.
Guidance and manufacturer declarations - electromagnetic
emissions
Emissions test
IEC60601
compliance
test level
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The Optos Table uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
NOT
likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Optos Table is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Pst
Dmax
Operations
The following procedures provide more information about using your table.
Adjusting the table height
WARNING
Ensure the cables are
NOT
pulled while the table height is being adjusted.
l
Select
to move the table up.
l
Select
to move the table down.
Contact us
Optos, Inc
Tel: 800-854-3039
[email protected]
Optos plc
Tel: +44 (0) 1383 843 300
[email protected]
Optos Australia
Tel: +61 8 8444 6500
[email protected]
Visit optos.com to find your local contact details.
Moving the table
The table has lockable wheels to prevent it moving during use.
1.
Remove all equipment from the table. Refer to the specific equipment instructions for details.
2.
Lower the height of the table to its lowest position for optimal stability and unobstructed view.
3.
Unplug the table mains cable from the wall socket.
4.
Place all cables on the table. Ensure there are no trailing cables.
5.
Pull the wheel lever up to unlock the wheel. Repeat for each wheel.
6.
Carefully move the table to its new location.
7.
When in position, push the wheel lever down to lock the wheel. Repeat for each wheel.
8.
Ensure the table works as expected.
General cleaning
CAUTION
Be careful
NOT
to damage the table or equipment when cleaning.
The table does not require cleaning before each use.
In a medical environment, if any part of the table or equipment becomes dirty or soiled with bodily fluids,
follow the general cleaning procedures and your internal clinic or hospital decontamination procedures.
Adjust the arm tension
To adjust the Monitor Support Arm tension, with a monitor fitted, follow these steps:
1.
Bend the main arm downwards to expose the bolt access cover.
2.
Remove and retain the access cover.
3.
Using an Hex key (4mm) turn the bolt as required.
a.
Turn the bolt clockwise to increase the tension and prevent the arm dropping.
b.
Turn the bolt anti-clockwise to decrease the tension and prevent the arm rising.
4.
Refit the access cover.
Adjust the monitor mount plate
To adjust the monitor mount plate tension follow these steps:
1.
Using an Hex key (4mm) turn the tension bolt either way as required:
a.
Turn "+" to increase tension.
b.
Turn "-" to decrease tension.