Procedure for Firmware upgrading
①
Save the firmware in any folder.
②
Select "Setting" on the main menu, and choose "Set Parameters" from the pulldown menu.
③
Click "Reference" to designate the folder where the firmware was saved in
①
.
④
Write the firmware name in "Inquiry:".
⑤
Click "Apply".
⑥
Click "OK".
⑦
Select "Run" on the main menu, and choose "Download" from the pulldown menu.