How to use Inspection Software
1. Purpose
This software is used when LS-5000ED is disassembled or its parts are replaced or inspections are made.
2. Preparation (tools required)
・
DOS/V PC
・
Inspection chart for
(
J61187
)
・
Positive film chart
(
J61189
)
; Insert this into MA-21
・
IR pass filter (J63088
)
; Insert this into MA-21
・
Inspection software (J65050A: Japanese, J65050B: English)
【
Hardware requirements
】
・
Windows2000
・
Japanese/English OS
・
CPU
Pentium
Ⅲ
700MHz or more
・
RAM
256MB or more
3. Inspection details
1) Initialization
・
Check the firmware version.
2) Illumination irregularity
・
Check for the raw data of the CCD with shading OFF in the standard range.
3) Shading
・
Check for the read-in data of the CCD waveform with shading ON in the standard range.
4) Noise inspection
・
By performing prescanning operation, check if the operating sound has no problem.
5) Mechanical precision
・
Scanning position accuracy
Get the central position (average values of x, y) of each alignment mark at 4 positions, then comparing them with set
values, check the deviation of image scanning position (shading).