MRT10*RDA001A/E
6 - 7
6.9 PI sensor boards
6.9.1 CON sensor
6.9.1.1 CON sensor of the upper brace R
Replacement part: 34085-BA08
1. Remove the covers. (See 5.1.)
2. Remove connector J9 on the MB main
board (34085-BA01).
3. Unscrew the tight screw P (82053-CE06B)
of the upper brace R (34085-2500) and
replace the PI sensor board (34085-BA08).
4. Reassemble the parts in reverse order.
6.9.1.2 CON sensor of the upper brace L
Replacement part: 34085-BA08
1. Remove the covers. (See 5.1.)
2. Remove connector J11 on the MB main
board (34085-BA01).
3. Unscrew the tight screw P (82053-CE06B)
of the upper brace L (34085-2600) and
replace the PI sensor board (34085-BA08).
4. Reassemble the parts in reverse order.
$#
%'$
%'$
$#
Summary of Contents for RT-5100
Page 6: ...1 2 MRT10 RDA001A E ...
Page 8: ...2 2 MRT10 RDA001A E ...
Page 10: ...3 2 MRT10 RDA001A E ...
Page 70: ...MRT10 RDA001A E 7 8 ...
Page 71: ...MRT10 RDA001A E 8 REFERENCES 8 1 8 1 Wiring diagram ...
Page 72: ...MRT10 RDA001A E 8 2 8 2 Connector cables 34085 EA01 34085 EA02 34085 EA04 34085 EA11 ...
Page 73: ...MRT10 RDA001A E 8 3 34085 EA21 ...