4 - 15
MRT10*RDA001A/E
4.3.4 “Chart doesn’t work.” appears.
Connect the chart projector.
Is the chart projector connected?
Is the power of the chart projector ON?
Does the communication cable break?
Do you try to change the chart while the chart is
moving?
It is recommended not to change the chart while the
chart is moving.
Turn ON the power of the chart projector.
Replace the communication cable.
Yes
No
No
No
Yes
No
Yes
Yes
Replace the RB I/F board (34085-BA22). (See 6.27.)
Is the symptom relieved?
Replace the RB main board (34085-BA21). (See 6.26.)
Yes
No
Is the chart projector applied to, or set for the RT-
5100?
Yes
Change the chart projector to the one applied to the
RT-5100, or change the setting of the chart projector to
the applied one.
No
Does cable CB break?
Replace cable CB.
No
Yes
Completion