3. OPERATING PROCEDURE : Bifocal and Trifocal Lens Measurement
3
When the target changes from
to
(cross), press the read button.
The distance power is locked in.
4
Measure the near add power (Add: 1st add
power).
Pull the lens forward to bring the near portion onto the
nosepiece.
When the add power is detected, the auto add measure-
ment function places the device into add power measure-
ment.
For trifocal lenses, bring the intermediate portion onto the
nosepiece.
5
Press the read button.
The measured data of the near add power (Add) is locked
in.
Because auto read does not function, press the read but-
ton to lock in the measured data.
When the measured prism value is not displayed, “Ad” is
displayed.
• In measurement of the distance portion, auto read works depending on the setting of the Auto read S
or Auto read R/L parameter.
• It is not necessary to align the target.
• Measure the near add power without removing the lens from the nosepiece.
Once the lens is removed, it is necessary to restart from the distance power measurement.
• For trifocal lenses, when the vertical length of the intermediate portion is narrow, the measuring beam
may interrupt the boundary and correct measurement may not be performed.
1707141668223 承認済
Uncontrolled copy when printed.
Summary of Contents for LM-7
Page 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Page 6: ...4 1707141668223 Uncontrolled copy when printed...
Page 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Page 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...