30
3. OPERATING PROCEDURE : Preparation for Measurement
♦
Measurement screen when power is turned on
Depending on the settings of the Initial screen and Contact measurement parameters and the type of
the attached nosepiece, the measurement screen displayed after initialization differs.
●
Initial screen parameter setting
●
Contact measurement parameter setting
●
When the parameter is set to “On” or “Off”
Except for the following cases, the measurement screen set by the Initial screen parameter is
displayed.
• When the Contact measurement parameter is set to “Off” and the Initial screen parameter is set to “Con-
tact”, the auto measurement screen is displayed.
• When the Initial screen parameter is set to any other than “Contact”, and power is turned on
with
the nosepiece for contact lenses mounted, the error message appears on the initial screen.
●
When the parameter is set to “Only”
Regardless of the Initial screen parameter setting,
the contact lens measurement screen is displayed.
Confirm that the nosepiece is for contact lenses.
Setting
Initial screen
Auto
Auto measurement screen
Normal
Normal measurement screen
Progressive
Progressive power lens measurement screen
Contact
Contact lens measurement screen
To measure spectacle lenses
Replace the nosepiece with the standard one, then press the restart
button .
The measurement screen set by the Initial screen parameter is dis-
played.
To measure contact lenses
only
Press the execute button
to display the contact lens mea-
surement screen.
In this case, the contact lens measurement screen cannot be
changed to the other measurement screens unless the nosepiece
is replaced and power is turned off and on.
1707141668223 承認済
Uncontrolled copy when printed.
Summary of Contents for LM-7
Page 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Page 6: ...4 1707141668223 Uncontrolled copy when printed...
Page 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Page 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...