103
6. SPECIFICATIONS AND TECHNICAL INFORMATION : EMC (Electromagnetic Compatibility)
* 1
For the regions where the rated voltage is 220 V or greater, this device complies with class A. For the regions where the
rated voltage is 127 V or less, this standard is not applicable.
* 2
For the regions where the rated voltage is 220 V or greater, this device complies with this standard. For the regions where
the rated voltage is 127 V or less, this standard is not applicable.
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
r
t
c
e
l
E
e
c
n
a
il
p
m
o
C
t
s
e
t
s
n
o
i
s
s
i
m
E
omagnetic environment - guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments, including domestic estab-
lishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
*1
Voltage fluctuations/ Flicker
emissions IEC 61000-3-3
*2
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic Dis-
charge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast tran-
sient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage, dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-
11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
) for 5 sec
<5%
U
T
(> 95% dip in
U
T
) for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
) for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
device requires continued operation during power
mains interruptions, it is recommended that the
device be powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
1707141668223 承認済
Uncontrolled copy when printed.
Summary of Contents for LM-7
Page 1: ...LM 7 LM 7P AUTO LENSMETER OPERATOR S MANUAL 1707141668223 Uncontrolled copy when printed...
Page 6: ...4 1707141668223 Uncontrolled copy when printed...
Page 12: ...10 1 SAFETY PRECAUTIONS Labels and Symbols 1707141668223 Uncontrolled copy when printed...
Page 88: ...86 4 CONNECTION AND SETTINGS Parameter Settings 1707141668223 Uncontrolled copy when printed...