25
DATA FORMAT
:
Example of Data Output
2
(10) In the case of the maximum data size in normal measurement mode (LM-1000/1000P)
Measurement conditions:
Trifocal lens measurement with the left- and right-eye lenses designated
Right: SPH = -1.25 D, CYL = -0.75 D, AXIS = 120º
PRISM = IN 2.25
Δ
, DOWN 2.00
Δ
(BU/D BI/O setting)
ADD = +2.00 D, ADD2 = +3.00 D
Representation form of near portion power = NEAR SPH
NSPH = +0.75D, NSPH2 = +1.75 D
Left: SPH = -2.00 D, CYL = -0.50 D, AXIS = 180º
PRISM = OUT 1.25
Δ
, UP 2.00
Δ
(BU/D BI/O setting)
ADD = +2.25 D, ADD2 = +3.50 D
Representation form of near portion power = NEAR SPH
NSPH = +0.25 D, NSPH2 = +1.50 D
CR code = Added
• This is an example of LM-1000P of the maximum data size (175 characters) in an ordinary
lens measurement mode.
• When the NEAR parameter is set to NEAR SPH, both the addition powers (ADD) and near
spherical data (NSPH) are output.
1
5
(Number of characters)
SOH
D
L
M
STX
→
To be continued
01H
44H 4CH 4DH
02H
10
15
20
I
D
N
I
D
E
K
/
L
M
-
1
0
0
0
P
ETB CR
→
To be
continued
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH
Company name/ Model name
25
30
35
40
R
-
0
1
.
2
5
-
0
0
.
7
5
1
2
0
ETB CR
→
To be
continued
20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH
Right SPH
Right CYL
Right AXIS
45
50
55
A
R
0
2
.
0
0
ETB CR
0
3
.
0
0
ETB CR
→
To be
continued
41H 52H 30H 32H 2EH 30H 30H 17H 0DH 30H 33H 2EH 30H 30H 17H 0DH
Right ADD
Right ADD2
60
65
70
75
N
R
+
0
0
.
7
5
ETB CR
+
0
1
.
7
5
ETB CR
→
To be
continued
4EH 52H 2BH 30H 30H 2EH 37H 35H 17H 0DH 2BH 30H 31H 2EH 37H 35H 17H 0DH
Right NSPH