DATA FORMAT
:
Example of Data Output
(8) In the case of the left and right data in progressive power lens measurement mode
Measurement conditions:
Progressive power lens measurement with the left- and right-eye lenses desig-
nated
Right:
SPH = -1.25 D, CYL = -0.75 D, AXIS = 120º
PRISM = IN 2.50
Δ
, DOWN 2.00
Δ
(BU/D BI/O setting)
ADD = +2.00 D
Left:
SPH = -2.00 D, CYL = -0.50 D, AXIS = 180º
PRISM = OUT 1.25
Δ
, UP 2.00
Δ
(BU/D BI/O setting)
ADD = +2.25 D
CR code = Added
• The addition power is added before the prism data.
• After the transmission of the right-eye lens data, the left-eye lens data is output.
SOH
D
L
M
STX
→
To be continued
01H
44H 4CH 4DH
02H
I
D
N
I
D
E
K
/
L
M
-
1
0
0
0
P
ETB CR
→
To be
continued
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH
Company name/ Model name
R
-
0
1
.
2
5
-
0
0
.
7
5
1
2
0
ETB CR
→
To be
continued
20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH
Right SPH
Right CYL
Right AXIS
A
R
0
2
.
0
0
ETB CR
P
R
0
2
.
5
0
I
ETB CR
→
To be
continued
41H 52H 30H 32H 2EH 30H 30H 17H 0DH 50H 52H 30H 32H 2EH 35H 30H 49H 17H 0DH
Right ADD
Right PRISM1
P
R
0
2
.
0
0
D
ETB CR
→
To be
continued
50H 52H 30H 32H 2EH 30H 30H 44H 17H 0DH
Right PRISM2
L
-
0
2
.
0
0
-
0
0
.
5
0
1
8
0
ETB CR
→
To be
continued
20H 4CH 2DH 30H 32H 2EH 30H 30H 2DH 30H 30H 2EH 35H 30H 31H 38H 30H 17H 0DH
Left SPH
Left CYL
Left AXIS
A
L
0
2
.
2
5
ETB CR
P
L
0
1
.
2
5
O
ETB CR
→
To be
continued
41H 4CH 30H 32H 2EH 32H 35H 17H 0DH 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH
Left ADD
Left PRISM1
P
L
0
2
.
0
0
U
ETB CR
EOT
1
6
C
1
CR
50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH
04H
31H 36H 43H 31H 0DH
Left PRISM2
Checksum