8
DATA FORMAT
:
Format of each data
2.2
Format of each data
The number inside the parentheses indicates the number of characters transmitted.
(1) Company name / Model name
The name of the manufacturer (NIDEK) and the name of model (LM-1000/LM-1000P/LM-
1200) are output.
* In the lower line, the ASCII codes are represented by hexadecimals.
• LM-1000P
• LM-1000
• LM-1200
(2) SPH, CYL, AXIS
Basic measured SPH, CYL, and AXIS data are output.
The number of digits of each measured data to be output is fixed with the leading digit 0
remaining.
The ETB and CR codes are added to the end (when the CR Code parameter is set to ON).
(The codes are added to the end of all the following types of data.)
I
D
N
I
D
E
K
/
L
M
-
1
0
0
0
P
ETB CR
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH
Company name
Model name (LM-1000P)
I
D
N
I
D
E
K
/
L
M
-
1
0
0
0
ETB CR
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 17H 0DH
Company name
Model name (LM-1000)
I
D
N
I
D
E
K
/
L
M
-
1
2
0
0
ETB CR
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 32H 30H 30H 17H 0DH
Company name
Model name (LM-1200)
Identification
code
Single
“ ” (Space, Space)
Right
“ R” (Space, R)
Left
“ L” (Space, L)
SPH
Four digits with a fixed decimal point
and a polarity sign
CYL
Four digits with a fixed decimal point
and a polarity sign
AXIS
Three-digit integer