background image

9

Operation

Pretest Checklist: 

Determine the size of the sample to be tested.  Adjust the 4 jig on the testing 
table accordingly.  Center the sample and use the knob to secure the sample in 
place (clockwise to tighten the jig to the sample material).  

It is important to make sure that all the jigs are flush to the moving brackets.  
Each one of these jigs has set pins which slides into place inside the grooved 
brackets.  

Measuring Modes  

Measuring Modes available on the TNP:  OPEN, CLOSE, MEAN 
Select the measuring mode required for testing.  Please check the settings below 
on how to select the right mode for testing.  

Pressing the Mode key toggles through the mode available from the TNP, LED 
indicators are provided to display current selected mode. 

OPEN and CLOSE modes are PEAK Values captured by the TNP, these values 
are not real time values, and are retained on the display until one of the following 
occurs, the Zero/RST key is pressed or a higher peak value is detected (which in 
this case, replaces the current value detected).  

Open mode 

Open 

Mean  

Close 

Mean mode 

Open 

Mean 

Close 

 Close mode 

Open 

Mean 

Close 

Nm 

Nm 

Nm 

Mode 

Mode 

Mode 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for TNP Series

Page 1: ...charge 16 hour max recharge time Operating Environment 32 104 F 0 40 C Dimensions 12 60 x 8 94 x 4 96 320 x 227x 126 mm Product Weight TNP 0 5 2 5 Approx 17 6 lb 8 kg TNP 10 27 5 lb 12 5 kg Package W...

Page 2: ...in parameter settings see next section Zero RST key Tare or zero function for resetting and initializing values while in Average or peak mode Secondary function in parameter settings see next section...

Page 3: ...before the unit turns off Main display Set up Values shows the measured value in four digits including decimal places While under function mode this displays the options for each setting Direction In...

Page 4: ...que meter is in Function Mode Operation key Operation How to operate Zero RST Power Function mode With Power off press and hold the Zero RST key then power ON the torque meter Continue holding the ZER...

Page 5: ...entation F04 0000 CCW Open 0000 CCW Open 0000 Upper comparator value Hi limit HI 0000 9999 with decimal point Setting the values to zero disables this function 0000 Lower comparator value Lo limit LO...

Page 6: ...ease the Power button but continue to hold Zero RST key until display F01 Function Mode Measurement unit Power Zero RST Unit N m Unit N cm Ncm Nm Unit Kg cm Kgc Unit lb in lbi TNP 2 5 10 Mode Mode Mod...

Page 7: ...n Mode Auto power OFF Disable auto power OFF Mode Mode Zero RST Pressing Zero RST it moves to the next function setting f04 Open f04 Function Mode Sign Orientation Open Mode Mode Zero RST Pressing Zer...

Page 8: ...all Recall Mem Mem Recall Zero RST Pressing Zero RST it moves to the next function setting Function Mode Lower comparator value Lo Limit Key operation is the same in entering the Hi limit value LO Pre...

Page 9: ...mode required for testing Please check the settings below on how to select the right mode for testing Pressing the Mode key toggles through the mode available from the TNP LED indicators are provided...

Page 10: ...O HI LO HI Red LED indicator will be lit on the Comparator LED indicators Utilizing this feature makes it an ideal tool for quality assurance checks Example HI is set to 100 and LO is set to 50 Based...

Page 11: ...F01 Press the Zero RST key until the main display shows HI limit Using the Recall key we can increment the highlighted digit from 0 9 Press the Mem key to move from left to right or to the next digit...

Page 12: ...ta can be stored in the TNP meter by pressing the Mem key This feature is only available in OPEN and CLOSE measuring modes NOTE In MEAN or Average mode the Mem key has no function When the number of d...

Page 13: ...es after pressing Recall To access the stored measured value use the Recall key to scroll thru the values This will be after the MIN value is displayed The display will flash two sets of numbers the f...

Page 14: ...X Maximum value CNT Alternating display 1sec interval Shows last memory data Returns to measuring mode Nm Nm Nm Alternating display 1sec interval Nm Nm Test number displayed Last memory number data di...

Page 15: ...e memory to be deleted jump out of sequence How does it work First select from Open or Close Modes Mean or average mode data cannot be stored Press the Recall key to access memory the first display yo...

Page 16: ...er off the TNP torque meter Press the Mem key and continue to hold this key while powering the torque tester The TNP will initialize itself You will see the model capacity displayed on the front panel...

Page 17: ...DIMENSIONS 17 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: