-13-
It's available to measure the compression and tension force. Measuring value appears at all times.
1. Press POWER key and release ( Turn POWER on after release)
2. Press ZERO key to tare.
Displayed measuring value is the averaged out sampling value (every 1 msec*) per display update time.
Display update time of default factory setting is 3 times/second. In order to increase the display response against the change of
measuring value, you may change the set value of display update time.
You can increase this time up to 20 times/second (regarding the change of display update time, please refer to Display Update
Time.)
To change the display units, just press UNIT and the units will change every time the button is pressed.
N; kg (g); lb (oz); N
Press ZERO key to reset the measuring value. Please press the ZERO key before starting the measurement in order not to change the
starting display value because of the own weight or measuring direction or weight of measuring fixture.
Measuring range is from maximum pulling load to maximum compression load. When measuring range is over the limit, “OVR” is
displayed.
Press the ZERO key under the plus peak hold mode or minus peak hold mode. The plus peak value or minus peak value is cleared. In
plus peak hold or minus peak hold modes, the tare cannot be performed even when pressing and releasing the ZERO key.
When turning the POWER on, a tare is automatically performed. If you turn POWER on while getting a load force, the display becomes
"0" and you cannot measure accurate values.
6.2.1Standard measuring mode
6.3.Change display unit
6.4.Tare
6.2.2 Peak hold mode
Peak mode displays teh greatest force in both positive or negative directions. Sampling time is 1ms.*
Press PEAK key to change standard measuring mode, Plus peak hold mode and Minus peak hold mode.
Under the plus peak hold mode, “PEAK” appears.
Under the minus peak hold mode, “PEAK” and “-” (minus) are displayed.
Key operation
Display
PEAK
ZERO
MEM
UNIT
Each time to press PEAK
Standard measuring mode
Plus peak hold mode
Minus peak hold mode
Under the plus peak hold mode and minus peak hold mode, press ZERO key to clear the peak value (Tare is not performed).
6.2.Measuring Mode
Standard measuring mode or peak hold modes are available.
Key operation
Display
PEAK
ZERO
MEM
UNIT
Press and release
Display force gauge model
Standard measuring mode
Display tracking ON/OFF
*Based on the Response Time (f05) setting.
*Based on the Response Time (f05) setting.
FGV