Nickel-Electro NE4-E Series Manual Download Page 8

The Clifton Range - 3 Year Warranty

8

High Quality  Products at Affordable Prices

Cleaning

General

Important - please follow these instructions to avoid possible damage to

the unit, otherwise affecting its performance and warranty.

Always disconnect the product from the electrical supply before

cleaning.

Cleaning External Painted Surfaces featuring “Anti-bacterial Paint Finish”

The water bath should be cleaned at regular intervals wiping external surfaces with a cloth or sponge
soaked in warm soapy water with a mild detergent. All surfaces should be cleaned using a soft cloth or
sponge.

Do not under any circumstances use strong solvents or solutions containi

ng Chlorinated Hydocarbons, Esters, Ketones or abrasive cleaners or polish

on the paint finish otherwise it damages the built

in anti-bacterial properties.

All painted surfaces on Clifton range products features an “Anti-bacterial
paint finish” identified with this authenticating logo on the unit.
This “Anti-bacterial paint finish” inhibits the growth of bacteria. It has been
tested by independent specialist test houses such as the Law Laboratories (in
the UK) using internationally recognized test methods and proven to be effec-
tive versus a wide range of bacteria species including Escherichia coli and
Staphylococcus aureus (MRSA).

We recognise hygienic coatings are part of a controlled approach to a cleaner
working environment. Within its formulation an active ingredient with proven anti-bacterial properties
is bound into the paint finish. The efficacy of the paint finish applied to the Clifton range is maintained
over its lifetime, as the anti-bacterial agent is integral within the paint.

In a laboratory environment it makes this one less source of contamination, contributing to essential
clean working practices. A benefit of such a paint finish can lead to a reduction in cleaning schedules
because surfaces are more protected and improves protection between cleaning. Unlike detergents
“Anti-bacterial paint finish” does not offer an instantaneous action, but is intended for long-term
general protection against bacterial growth.

Moisture on the painted surface is necessary for the bacterium to absorb the agent and be affected
by it. The coating is therefore less active in very dry conditions, but dependent upon relative humid-
ity, moisture in the atmosphere maintains activity. Areas where moisture is trapped are also areas
that normally are difficult to clean and where bacteria proliferate but these areas are most active for
the anti-bacterial coating improving the defence against bacterial growth.

Cleaning the Stainless Steel Tank

The stainless steel crevice free tank with smooth corners should provide years of valuable service and is
resistant to chloride containing solutions it is however important to avoid high concentrations of halo-
gens - especially chloride. With such a high quality and resistant tank it may show symptoms of these
halogens as rust, which are deposits from external sources in the water supply.

We recommend always empty the bath of liquid after use and wipe out the internal faces of the tank with
a non-abrasive cloth and allow to dry. Any deposits can be removed with nitric acid (10%) on a cloth.

WEAR PROTECTIVE EQUIPMENT!

It is also recommended to use an accessory lid to prevent contaminates landing in bath liquids.

Summary of Contents for NE4-E Series

Page 1: ...level float switch and if the liquid drops below the recommended level heating is automatically switched OFF top up liquid level then switch off and back on again using the mains switch located on th...

Page 2: ...painted surfaces with Anti bacterial finish 8 Cleaning the stainless steel tank 8 Descaling the stainless steel tank 9 Decontaminationof Equipment 9 VirucidalDisinfectant 9 Disinfectant Sterilant 9 3...

Page 3: ...olypropylene spheres when operating at temperatures above 60 C Take care when lifting the lid Steam and hot vapours can cause scalding 13 Always use the display or a thermometer to check the temperatu...

Page 4: ...rs supplied Turn the thermostirrer upside down and screw the nylon propellor hand tight onto the spindle The spindle is threaded in the opposite direction to the propellor to ensure the propellor stay...

Page 5: ...electrical supply Suitable Liquids Operating temperatures and recommended liquid options Ambient to 99 C Distilled water Heat transfer liquid The LB range is formulated for temperatures from 45 C to 9...

Page 6: ...ater bath NE4 T series stainless steel tank with false base shelf fitted inside NE4 E series clear tank with shelf fitted Indicators heating temperature alarm timer and set temperature Tank View Low l...

Page 7: ...to stabilise and take an actual reading at the far end of the bath to the Thermostirrer Where necessary adjust the temperature controller Heater indicator will illuminate Low Liquid Level Float Switc...

Page 8: ...nge is maintained over its lifetime as the anti bacterial agent is integral within the paint In a laboratory environment it makes this one less source of contamination contributing to essential clean...

Page 9: ...ment and instruments where autoclaving is neither practical nor necessary For more detailed information relat ing to how Virkon should be used with access to test reports www relyon dupont com Is Virk...

Page 10: ...lifton prod ucts are easily recycled with majority of the product constructed from stainless or mild steels which can readily be re used or recycled In excess of 78 of this product range can be easily...

Page 11: ...to suit 22 and 28 Litre Baths Stainless Steel Test Tube Racks Dimensions 270 x 70 x 138mm L x W x H 6870 Stainless Steel Test Tube Rack 26 Hole x 17mm Diameter 6871 Stainless Steel Test Tube Rack 16 H...

Page 12: ...Place the loaded test tube rack into the bath on top of either the stainless steel perforated shelf or the raised shelf Thermometer Clip This can be positioned anywhere on the perimeter of the bath b...

Page 13: ...Spares Service Diagram The Clifton Range 3 Year Warranty 13 High Quality Products at Affordable Prices...

Page 14: ...t finish and is complete with a stainless steel shelf designed for good temperature stability STIRRED Features Include Sensitivity 0 2 C Uniformity 0 05 C Temperature range ambient 5 C 99 C Thermostat...

Page 15: ...for electrical equipment for measurement control and laboratory use BS EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements NICKEL ELECTRO LTD Manufacturers of lab...

Page 16: ...BX0844 BX0845 www nickel electro co uk...

Reviews: