Summary of Contents for NE4-E Series

Page 1: ...level float switch and if the liquid drops below the recommended level heating is automatically switched OFF top up liquid level then switch off and back on again using the mains switch located on th...

Page 2: ...painted surfaces with Anti bacterial finish 8 Cleaning the stainless steel tank 8 Descaling the stainless steel tank 9 Decontaminationof Equipment 9 VirucidalDisinfectant 9 Disinfectant Sterilant 9 3...

Page 3: ...olypropylene spheres when operating at temperatures above 60 C Take care when lifting the lid Steam and hot vapours can cause scalding 13 Always use the display or a thermometer to check the temperatu...

Page 4: ...rs supplied Turn the thermostirrer upside down and screw the nylon propellor hand tight onto the spindle The spindle is threaded in the opposite direction to the propellor to ensure the propellor stay...

Page 5: ...electrical supply Suitable Liquids Operating temperatures and recommended liquid options Ambient to 99 C Distilled water Heat transfer liquid The LB range is formulated for temperatures from 45 C to 9...

Page 6: ...ater bath NE4 T series stainless steel tank with false base shelf fitted inside NE4 E series clear tank with shelf fitted Indicators heating temperature alarm timer and set temperature Tank View Low l...

Page 7: ...to stabilise and take an actual reading at the far end of the bath to the Thermostirrer Where necessary adjust the temperature controller Heater indicator will illuminate Low Liquid Level Float Switc...

Page 8: ...nge is maintained over its lifetime as the anti bacterial agent is integral within the paint In a laboratory environment it makes this one less source of contamination contributing to essential clean...

Page 9: ...ment and instruments where autoclaving is neither practical nor necessary For more detailed information relat ing to how Virkon should be used with access to test reports www relyon dupont com Is Virk...

Page 10: ...lifton prod ucts are easily recycled with majority of the product constructed from stainless or mild steels which can readily be re used or recycled In excess of 78 of this product range can be easily...

Page 11: ...to suit 22 and 28 Litre Baths Stainless Steel Test Tube Racks Dimensions 270 x 70 x 138mm L x W x H 6870 Stainless Steel Test Tube Rack 26 Hole x 17mm Diameter 6871 Stainless Steel Test Tube Rack 16 H...

Page 12: ...Place the loaded test tube rack into the bath on top of either the stainless steel perforated shelf or the raised shelf Thermometer Clip This can be positioned anywhere on the perimeter of the bath b...

Page 13: ...Spares Service Diagram The Clifton Range 3 Year Warranty 13 High Quality Products at Affordable Prices...

Page 14: ...t finish and is complete with a stainless steel shelf designed for good temperature stability STIRRED Features Include Sensitivity 0 2 C Uniformity 0 05 C Temperature range ambient 5 C 99 C Thermostat...

Page 15: ...for electrical equipment for measurement control and laboratory use BS EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements NICKEL ELECTRO LTD Manufacturers of lab...

Page 16: ...BX0844 BX0845 www nickel electro co uk...