3.3 Monitor and control signal system
HVA4321
3-11
3.3 Monitor and control signal system
3.3.1 Monitor
MONITOR VOLTAGE
It is the BNC connector for the monitor signal of output voltage.
・
Gain: 1V/1kV
・
Output impedance: 50
Ω±
10%
MONITOR CURRENT
It is the BNC connector for the monitor signal of output current. In constant current mode, current
selected by TOTAL/LOAD is monitored.
・
Gain: 1V/10 mA
・
Output impedance: 50
Ω±
10%
Figure3-15. Monitor output
SIGNAL ON/OFF STSTUS
It is the total level signal output that indicates SIGNAL ON/OFF status. When the signal is ON, the
level is low (approx. 0.1V or less). When the signal is OFF, the level is high (approx. 4.5V or more).
1/ Signal switch
Figure3-16. Status signal output
SI GNAL ON/ OFF STATUS
OUTPUT
MONI TOR
VOLTAGE
CURRENT
TTL(Lo: ON)
Zout 10k
Ω
1V/ 1kV
Zout 50
Ω
1V/ 10mA
Zout 50
Ω
SI GNAL
ス イ ッ チ
SI GNAL ON/OFF
STATUS
ON
OFF
LOW
HI GH
HI GH
OFF
SIGNAL SWITCH
SIGNAL ON/OFF
STATUS
Summary of Contents for HVA4321
Page 1: ...INSTRUCTION MANUAL NF Corporation HVA4321 10kV AC DC AMPLIFIER HVA4321 10kV AC DC AMPLIFIER...
Page 2: ......
Page 4: ......
Page 8: ...HVA4321 iv...
Page 16: ...Figures HVA4321 xii...
Page 18: ...HVA4321 1 2...
Page 24: ......
Page 30: ...2 3 Grounding and power supply connection HVA4321 2 6...
Page 32: ......
Page 50: ......
Page 52: ......
Page 54: ......
Page 56: ......
Page 116: ......
Page 127: ...6 Troubleshooting 6 1 When suspecting a fault 6 1...
Page 128: ......
Page 132: ......
Page 138: ......
Page 148: ......
Page 150: ......
Page 151: ......